High pressure steam sterilizer
is a kind of equipment that uses saturated steam to quickly and
reliably sterilize or sterilize articles. It is widely used in medical
and health systems, biotechnology, food and drug production and
processing. Sterilization and sterilization of medical devices,
dressings, glassware, and solution culture media also apply to
high-altitude drinking water in highland areas for cooking equipment and
enterprises and institutions, as well as high-temperature steam source
equipment.
Sterilization time of autoclave:
1.
Heat penetration time is the time required to reach the sterilization
temperature from the sterilization temperature in the disinfecting
cabinet to the center of the disinfected article. The length of time
depends on the nature of the disinfectant, the size of the package, the
location and the degree of air emptying in the autoclave and the type of
sterilizer.
2, disinfection and maintenance
time, that is, the time required to kill microorganisms, generally used
to kill spores of T. stearothermophilus spores required time to express.
It takes 30 minutes at 115°C, 12 minutes at 121°C, and 2 minutes at 132°C.
3,
safety time, generally half of the maintenance time, depending on the
length of the disinfection items. Sterilization of easily thermally
conductive metal devices requires no safety time.